Wallmounted Oxygen Flowmeter
| Product Code | Type |
|---|---|
| FLOW-O2-15 | Flow Meter Oxygen – 15l/min & Humidifier |
| HUMIDIFIER | Humidifier Bottle with Lid – 150ml |
₨1.00
Add to cart
CPX DIN Medical Gas Outlets are manufactured and designed in our UK factory, with full metal construction, Gas Indexing as EN 13260-2, EN 9170, and Medical Gas Pipelines EN 7396, DIN 13260-2 Part 2: Dimensions and allocation of probes and gas-specific connection points for Terminal Units. CPX DIN Gas Outlets are available for wall mounting (Flush or Surface), Bed Head trunking, and Pendant mounting. Can be supplied with metal or plastic second fix.
Gas Specificity Flowmeters shall be supplied with gas-specific probes and only attach to the corresponding Medical Gas Terminal Unit, it accurately measures flow at 0-15Lpm even at variant inlet pressures. Accuracy is fulfilled within 10% to attain maximum efficiency for the patient and environment. Each flow meter is calibrated at 50 psi inlet pressure Materials The Flowmeter shall be manufactured to EN 13220:1999 and ISO 15002:2008 and constructed from a chromium-plated brass body, inner and external tube in crack-resistant polycarbonate and stainless steel ball. Other pieces shall be in brass. Humidifier: Constructed from polycarbonate, sterilisation should be conducted in an autoclave on a 25 min cycle at 121 ºC, 1.3 kPa (max. 40 times). Before sterilisation remove the tube and diffuser and consequently disposed. Quality: Manufactured under BS EN 13485 Medical Devices: Quality Management Systems.
| Product Code | Type |
|---|---|
| FLOW-O2-15 | Flow Meter Oxygen – 15l/min & Humidifier |
| HUMIDIFIER | Humidifier Bottle with Lid – 150ml |